Mc-techgroup SP2000H_DIL_BR Operator's manual Manual do Utilizador Página 10

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 35
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 9
10
Gas sampling and gas conditioning technology 2-1.1.7-ME
Sample out
diluted
Pressure gauge
-1…0bar
Cal gas
Dil gas
Check valve
Pre-heater
Filter
Filter
Crit. orifice
Bypass back to
The process
Bypass gas
View "A"
A
A
Figure 3 Probe type SP2000H/DIL/BR
To prevent cooling below dew point at the dilution point, the dilution unit with critical orifice is in-
stalled temperature-stable in the heated part of the gas sample probe directly in the clean gas outlet.
A gas pre-heater heats the dilution gas to probe temperature. With the option ..B/..BR wit bypass in-
jector the bypass gas is pre-heated as well (double pre-heater). The probe is available with heating to
maximum 180°C or 320°C.
Test gas can be supplied to the probe for analyser calibration via the integrated test gas feed connec-
tion .
Vista de página 9
1 2 ... 5 6 7 8 9 10 11 12 13 14 15 ... 34 35

Comentários a estes Manuais

Sem comentários