Mc-techgroup EC-Ex Data sheet Manual do Utilizador Página 2

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 4
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 1
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
6.5
Technical specications and illustrations are without
obligation, subject to modications. 10.96/09.09
Dimensions in mm
Gas cooler ECS...
435
85
360
205
465
483
50
235
310
Gas Inlet/outlet
Condensate
EC
Cooler
C <
C >
ON
Ø70
Application example for ECS
1 Heated lter sample probe SP210-H or SP2000-H
2 Heated sample line 4M4/6
3 Cooler ECS
4 3-way ball valve 3L/PV-1
5 Peristaltic pump SR25.2
6 Diaphragm pump MP47 or MP06/12 or N5KP
7 Fine lter FP-2T-D with liquid alarm LA1
8 Aerosol lter CLF-5 /W optional according to application
9 Flow meter FM10 or FM40, 25-250 Nl/hr
10 Analysers f. e. PMA100
8
Alternative:
trap with float or condensate
collection vessel
5
condensate OUT
+5°C
4
3
10
9
7
ECS
1 2
sample gas IN
test gas
IN
6
Dimensions
Cooler type EC-SS with heat exchanger out
of SS316Ti
Cooler type EC-PV with haet exchanger out
of PVDF
Gas flow rate
10
8
6
4
10 30 50 70 90 110 130 150 170 190 210 230 250 Nl/h
°C
Sample outlet dew point
Sample inlet dew point
60 °C
40 °C
Ambient temp. 20 °C
10
8
6
4
10 30 50 70 90 110 130 150 170 190 210 230 250 Nl/h
°C
Gas flow rate
Sample outlet dew point
Sample inlet dew point
Ambient temp. 20 °C
60 °C
40 °C
10
8
6
4
10 30 50 70 90 110 130 150 170 190 210 230 250 Nl/h
°C
Gas flow rate
Sample outlet dew point
Sample inlet dew point
Ambient temp. 20 °C
60 °C
40 °C
Cooler type EC-G with heat exchanger out of
glass
Vista de página 1
1 2 3 4

Comentários a estes Manuais