10 Gas sampling and gas conditioning technology 14.1a-ME
8 TECHNICAL SPECIFICATIONS
Gas conditioning series CSS ®
Article number 115V/50-60Hz
Setting range: + 2 ° C ..... + 7 ° C, factory setting: + 5 ° C
under constant conditions < ± 0.1 ° C
Gas inlet water-vapour saturation
Total cooling capacity **
Material of parts in contact with
medium
Heat exchangers: Glass, PVDF or stainless steel 1.4571
Hoses/screwed connectors: PTFE/PVDF
Universal filter e.g. FP-2T-D: PVDF, FPM, glass, PTFE
Front-installed filter FPF-0.1GF: PVDF, glass, FPM
Front-installed filter FPF-2-0.3GF: PVC, FPM, PPH
Peristaltic pump SR25.2: PVDF, Novoprene ®
Liquid particle e.g. CLF-5/W: PVDF, glass, FPM, PTFE
Sample gas pumps N3/5/9: PVDF, PTFE, FPM
Flow meter FM40: PVDF, glass, FPM, Hastelloy C4
Ball valve 3L/PV and 5L/PV: PVDF, FPM
230V 50-60Hz ± 10% or 115V 50-60Hz ± 10% ***
max. 220VA + max. 300VA for sample gas pumps
- Cold-device connector with 2 m cable
- Single alarm connections (11 & 12-pole) or status alarm(s)
(6-pole) : maximum capacity of the relays 24V 500mA
- mA-outlet : maximum load 500Ohm, accuracy 2% of
measuring range end value for display indication
- M&C Bus
- Heated sample line : maximum length 10m
Steel housing for 19"- or wall mounting, painted RAL 7035
Device dimensions (H x W x D)
267.5 mm x 483 mm x 301.5 mm
(Portable version: 355 mm x 515 mm x 395 mm)
Electrical device standard
FPM = Viton ®
PVDF = Polyvinylidene fluoride
PVC = Polyvinyl chloride
FPM = Fluorinated rubber
PPH = Hard polypropylene
PTFE = Polytetrafluoroethylene
**Technical specifications with maximum values must be rated in consideration of total cooling capacity at 25° C and an outlet dew point of 5 ° C.
*** 115V/50Hz sample gas pump available as a special item